Módulo de Registro de EEG con Electrodos Activos para Interfaz Cerebro-Computadora basada en Imaginación Motora (Embedded EEG Recording Module with Active Electrodes for Motor Imagery Brain-Computer Interface)

Jorge Airy Mercado (mercado.jorgea@gmail.com)1, Javier Herrera (javier.herrera_91@hotmail.com)2, Arturo de Jesús Pansza (pansza-cyborg@gmail.com)3, Josefina Gutiérrez (josefina_gutierrez@hotmail.com)1


1Instituto Nacional de Rehabilitación
2Universidad La Salle
3Instituto Politécnico Nacional

This paper appears in: Revista IEEE América Latina

Publication Date: Feb. 2016
Volume: 14,   Issue: 2 
ISSN: 1548-0992


Abstract:
This paper presents a novel modular, portable and low-power electroencephalography (EEG) acquisition system for a Brain-Computer Interface (BCI) application. The system is based on the versatile microcontroller MSP430F5529, to acquire from 4 to 12 EEG channels at 256 Hz with a single 3.7 V supply. The prototype supports both passive and active electrodes; these were also designed and built. Additionally, a graphic visualization interface was developed on the open-source programming language Processing. Common-mode rejection ratio, input-referred noise and magnitude and phase frequency response were measured for each analog EEG channel, these requirements comply the International Federation of Clinical Neurophysiology guidelines. Furthermore, four noise and distortion parameters for AC input signals were evaluated for each ADC channel, based on the IEEE-1241-2000 standard. Power spectrum of EEG recordings using our prototype and the commercial amplifier g.USBamp were compared in open/closed eyes conditions (alfa-band reactivity). It was verified that spectral information retrieval (EEG alfa peak) attained with the prototype is quite similar to the commercial system, and that 60 Hz noise level is reduced when using active instead of passive electrodes, as expected. It was shown that analog pre-processing and noise design considerations are effective ways to get a good quality EEG signal and to enhance certain portions of its frequency spectrum (0.1‑35 Hz). This embedded recording system improves the acquisition of EEG into the frequency bands of interest; the mu (8-13 Hz) and beta (14-27 Hz) EEG rhythms for a motor-imagery BCI application, while rejecting unwanted components (DC, >30 Hz, 60 Hz). With this design, it is possible to reduce noise and interferences in EEG signal to decrease the computational workload in digital post-processing, and thus, may increase the likelihood of achieving a completely stand-alone BCI system for clinical applications.

Index Terms:
EEG, BCI, Active Electrodes, IEEE-1241-2000, IFCN   


Documents that cite this document
This function is not implemented yet.


[PDF Full-Text (825)]