|
Superconducting Spectrometer Magnet, G0
Taking Shape at BWX Technologies
Lynchburg, VA-- A large toroidal superconducting magnet system is being
assembled at the Lynchburg Magnet Facility of BWX Technologies. This magnet is
slated for recoil spectrometer of a Photonuclear Physics experiment to be
performed at Jefferson Lab in 2001. BWXT is under contract to the University of
Illinois http://www.npl.uiuc.edu/exp/G0/G0Main.html to design and build the
magnet system for this experiment. The magnet consists of 8 large
superconducting coils arranged around the axis of a stainless steel and aluminum
vacuum vessel, as shown in the scale model below.

The magnet system has a number of significant and novel features. BWXT was
provided with the required functional specification for the cold excited current
distribution in space, and had to back out a warm as constructed design that
would meet those requirements. The resulting warm design has very tight
tolerances on the conductor thickness, turn spacing, and coil shape that
required detailed warm mapping of the field for verification prior to the start
of the cold mass assembly. The coils have been completed and assembled, as shown
below.

The cryo-cooled production target for the experiment is to be located at the
center of the magnet. In order to sort the desired scattered particles, several
tons of lead collimators block mounted on aluminum frames are interleaved with
the coils. In the operational orientation, these collimators are clocked at all
angles, resulting in complex static loads that must be compensated to preserve
the sector field symmetry. The large attractive forces (ala tokamak toriodal
coils) must in addition be bucked at the center. Extensive coupled
structural-thermal and structural-electromagnet calculations were performed to
validate the design basis during the final design process. This analysis
included the conductor and turn, layer and ground insulation in addition to all
structural components. The magnet was solid modeled from that resulting analysis
basis and fabrication drawings created directly from the 3D solid models.

All of the coil structure is made of aluminum in order to preserve the
symmetry of the distribution of current in space. The coils are conductivity
cooled through internal channels bored in the aluminum bobbins. Cryogens are
circulated in 4 thermal-siphon loops of two coils each. The coils are composed
of two double pancakes on roughly rectangular bobbins. In general, conductors
were place to within 2-3 mils of their intended locations during winding.
The field symmetry requirements of the eight magnetic sectors are such that
vessel materials and construction were also critical to the design solution.
Materials with magnetic permeability greater than 1.003 were prohibited in the
vessel end caps- resulting in structural aluminum as the construction material
of choice, but difficult given the large size and required flat planar geometry.
Materials with magnetic permeability greater that 1.002 were prohibited for use
the vessel shell, but envelope constraints required minimum overall size. This
was solved by developing a special stainless steel chemistry and welding
techniques, that resulted the 42 foot circumference vessel having peak magnetic
permeability's not greater that 1.007-1.009.

The final assembly of the magnet is now in progress and acceptance testing
will be performed at UIUC in September. Final shipping to Jefferson Lab should
occur in early 2001. For more information contact:
Dr. Timothy A. Antaya
G0 Project manager
BWX Technologies
Lynchburg, VA, 24551
(804) 522 5565
timothy.a.antaya@mcdermott.com
|